Remote control commands
R&S
®
SMA100B
508User Manual 1178.3834.02 ─ 09
:SENSe<ch>[:POWer]:SWEep:TIME[:SENSor]:PULSe:STATe <State>
Enables pulse data analysis. The measurement is started with command INITiate.
Note: The command is only available in time measurement mode and with R&S NRP-
Z81 power sensors.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
SENS1:SWE:TIM:PULS:STAT ON
Enables pulse data analysis.
Options: Option R&S SMAB-K28
Manual operation: See "Pulse Analysis" on page 266
:SENSe<ch>[:POWer]:SWEep:TIME[:SENSor]:PULSe:THReshold:BASE
<Base>
Selects how the threshold parameters for pulse analysis are calculated.
Note: The command is only available in time measurement mode and with R&S
NRPZ81 power sensors.
Parameters:
<Base> VOLTage | POWer
*RST: VOLTage
Example:
SENS1:SWE:TIME:PULS:THR:BASE POW
Activates threshold calculation related to power.
Options: Option R&S SMAB-K28
Manual operation: See "Base" on page 266
:SENSe<ch>[:POWer]:SWEep:TIME[:SENSor]:PULSe:THReshold:POWer:
HREFerence <HReference>
Sets the upper reference level in terms of percentage of the overall pulse level (power
or voltage). The distal power defines the end of the rising edge and the start of the fall-
ing edge of the pulse.
Note: The command is only available in time measurement mode and with R&S
NRPZ81 power sensors.
Parameters:
<HReference> float
Range: 0 to 100
Increment: 0.01
*RST: 90
Options: Option R&S SMAB-K28
Manual operation: See "Distal" on page 266