:SENSe<ch>[:POWer]:SWEep:POWer[:SENSor]:SFRequency:STATe <State>
Activates the use of a different frequency for the power measurement.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
SENS1:SWE:POW:SENS:SFR:STAT ON
Activates the use of a separate frequency than the generator fre-
quency for power analysis
Manual operation: See "Use Separate Frequency- Power Analysis" on page 261
:SENSe<ch>[:POWer]:SWEep:TIME[:SENSor]:OFFSet <Offset>
Defines the level offset at the sensor input in dB. Activate the offset with the com-
mand :SENSe<ch>[:POWer]:SWEep:TIME[:SENSor]:OFFSet:STATe
on page 507.
Parameters:
<Offset> float
Range: -100 to 100
Increment: 0.01
*RST: 0
Example:
SENS2:SWE:TIME:OFFS -7dB
Defines a level offset of -7 dB.
SENS2:SWE:TIME:OFFS:STAT ON
The specified level offset is considered.
Options: Option R&S SMAB-K28
Manual operation: See "Level Offset - Power Analysis" on page 261
:SENSe<ch>[:POWer]:SWEep:TIME[:SENSor]:OFFSet:STATe <State>
Activates a level offset at the sensor input. Define the appropriate value with the com-
mand :SENSe<ch>[:POWer]:SWEep:TIME[:SENSor]:OFFSet on page 507.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
SENS2:SWE:POW:TIME:STAT ON
Activates the specified level offset.
Options: Option R&S SMAB-K28
Manual operation: See "Level Offset State- Power Analysis" on page 261