Remote control commands
R&S
®
SMA100B
506User Manual 1178.3834.02 ─ 09
Manual operation:  See "Use Separate Frequency- Power Analysis" on page 261
:SENSe<ch>[:POWer]:SWEep:POWer[:SENSor]:OFFSet <Offset>
Defines the level offset at the sensor input in dB. Activate the offset with the com-
mand :SENSe<ch>[:POWer]:SWEep:POWer[:SENSor]:OFFSet:STATe
on page 506.
Parameters:
<Offset> float
Range:  -100  to  100
Increment:  0.01
*RST:  0
Example: 
SENS2:SWE:POW:OFFS -5dB
Defines a level offset of -5 dB.
SENS2:SWE:POW:OFFS:STAT ON
Activates that the specified level offset is considered.
Options:  Option R&S SMAB-K28
Manual operation:  See "Level Offset - Power Analysis" on page 261
:SENSe<ch>[:POWer]:SWEep:POWer[:SENSor]:OFFSet:STATe <State>
Activates a level offset at the sensor input. Define the appropriate value with the com-
mand :SENSe<ch>[:POWer]:SWEep:POWer[:SENSor]:OFFSet on page 506.
Parameters:
<State> 0 | 1 | OFF | ON
*RST:  0
Example: 
SENS2:SWE:POW:OFFS:STAT ON
Activates the specified level offset.
Options:  Option R&S SMAB-K28
:SENSe<ch>[:POWer]:SWEep:POWer[:SENSor]:SFRequency <SFRequency>
Defines the separate frequency used for power vs. power measurement.
Parameters:
<SFRequency> float
Range:  0  to  1E12
Increment:  1
*RST:  1E6
Example: 
SENS1:SWE:POW:SENS:SFR 2GHz
The measurement is performed at 2 GHz
Manual operation:  See "Use Separate Frequency- Power Analysis" on page 261
Power sensor measurement subsystems