Remote control commands
R&S
®
SMA100B
505User Manual 1178.3834.02 ─ 09
Example:
SENS2:SWE:FREQ:SENS:SRAN:STAT ON
Activates use of a separate frequency range for frequency ver-
sus power measurement for sensor 2.
SENS2:SWE:FREQ:SENS:STAR 2.0GHZ
Sets a sweep start at 2 GHz irrespective of the current signal
generator frequency settings.
Manual operation: See "Min Frequency - Power Analysis" on page 262
:SENSe<ch>[:POWer]:SWEep:FREQuency[:SENSor]:SRANge:STOP <Stop>
Sets the stop frequency for the frequency power analysis with separate frequencies.
Parameters:
<Stop> integer
Range: 0 to 1E12
*RST: 10E6
Example:
SENS:SWE:FREQ:SENS2:SRAN:STAT ON
Activates use of a separate frequency range for frequency ver-
sus power measurement.
SENS:SWE:FREQ:SENS2:STAR 2.0GHZ
Sets a sweep start at 2 GHz irrespective of the current signal
generator frequency settings.
SENS:SWE:FREQ:SENS2:STOP 2.9GHZ
Sets a sweep stop at 2.9 GHz irrespective of the current signal
generator frequency settings.
Manual operation: See "Max Frequency - Power Analysis" on page 262
:SENSe<ch>[:POWer]:SWEep:FREQuency[:SENSor]:SRANge[:STATe] <State>
Activates the use of a frequency range for the power measurement that is different to
the set signal generator frequency range. The separate frequency range is entered
with commands :SENSe<ch>[:POWer]:SWEep:FREQuency[:SENSor]:SRANge:
STARt on page 504 and :SENSe<ch>[:POWer]:SWEep:FREQuency[:SENSor]:
SRANge:STOP on page 505.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
SENS2:SWE:FREQ:SENS:SRAN ON
Activates use of a separate frequency range for frequency ver-
sus power measurement for sensor 2.
SENS2:SWE:FREQ:SENS:STAR 2.0GHZ
Sets a sweep start at 2 GHz irrespective of the current signal
generator frequency settings.
SENS2:SWE:FREQ:SENS:STOP 2.9GHZ
Sets a sweep stop at 2.9 GHz irrespective of the current signal
generator frequency settings.
Power sensor measurement subsystems