Remote control commands
R&S
®
SMA100B
504User Manual 1178.3834.02 ─ 09
:SENSe[:POWer]:SWEep:TIME:YSCale:AUTO................................................................. 532
:SENSe[:POWer]:SWEep:TIME:YSCale:AUTO:RESet.......................................................533
:SENSe[:POWer]:SWEep:TIME:YSCale:MAXimum........................................................... 533
:SENSe[:POWer]:SWEep:TIME:YSCale:MINimum............................................................ 534
:SENSe<ch>[:POWer]:SWEep:FREQuency[:SENSor]:OFFSet <Offset>
Defines the level offset at the sensor input in dB. Activate the offset with the com-
mand :SENSe<ch>[:POWer]:SWEep:FREQuency[:SENSor]:OFFSet:STATe
on page 504.
Parameters:
<Offset> float
Range: -100 to 100
Increment: 0.01
*RST: 0
Example:
SENS2:SWE:FREQ:OFFS -3dB
Defines a level offset of -3 dB.
SENS2:SWE:FREQ:OFFS:STAT ON
Activates the specified level offset.
Options: Option R&S SMAB-K28
Manual operation: See "Level Offset - Power Analysis" on page 261
:SENSe<ch>[:POWer]:SWEep:FREQuency[:SENSor]:OFFSet:STATe <State>
Activates a level offset at the sensor input. Define the appropriate value with the com-
mand :SENSe<ch>[:POWer]:SWEep:FREQuency[:SENSor]:OFFSet
on page 504.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
SENS2:SWE:FREQ:OFFS:STAT ON
Activates the specified level offset.
Options: Option R&S SMAB-K28
Manual operation: See "Level Offset State- Power Analysis" on page 261
:SENSe<ch>[:POWer]:SWEep:FREQuency[:SENSor]:SRANge:STARt <Start>
Sets the start frequency for the frequency power analysis with separate frequencies.
Parameters:
<Start> integer
Range: 0 to 1E12
*RST: 1E6
Power sensor measurement subsystems