Remote control commands
R&S
®
SMA100B
514User Manual 1178.3834.02 ─ 09
:SENSe[:POWer]:SWEep:FREQuency:STARt <Start>
Sets the start frequency for the frequency mode.
Parameters:
<Start> float
Range: 0 to 1E12
*RST: 1E6
Example:
SENS:SWE:FREQ:STAR 2.0GHZ
Sets a sweep start at 2 GHz.
Manual operation: See "Min - Power Analysis" on page 256
:SENSe[:POWer]:SWEep:FREQuency:STEPs <Steps>
Sets the number of measurement steps for the frequency mode.
Parameters:
<Steps> integer
Range: 10 to 1000
*RST: 200
Example:
SENS:SWE:FREQ:STEP 500
Sets 500 steps
Manual operation: See "Steps - Power Analysis" on page 257
:SENSe[:POWer]:SWEep:FREQuency:STOP <Stop>
Sets the stop frequency for the frequency mode.
Parameters:
<Stop> float
Range: 0 to 1E12
*RST: 22GHZ
Example:
SENS:SWE:FREQ:STOP 20.0GHZ
Sets the sweep stop to 20 GHz
Manual operation: See "Max - Power Analysis" on page 257
:SENSe[:POWer]:SWEep:FREQuency:TIMing[:MODE] <Mode>
Selects the mode in terms of speed and precision of the response of a measurement.
Parameters:
<Mode> FAST | NORMal | HPRecision | FAST | NORMal
FAST
Selection FAST leads to a fast measurement with a short inte-
gration time for each measurement step.
Power sensor measurement subsystems