Remote control commands
R&S
®
SMA100B
536User Manual 1178.3834.02 ─ 09
:TRACe<ch>:FREQ:SWEep:SRC <FreqSource>
Determines the source of a trace for display in frequency measurement mode.
Parameters:
<FreqSource> OFF | SEN1 | SEN2 | SEN3 | SEN4 | HOLD | REF | ON
ON|OFF
Activates ofr deactivates the display of a trace.
SEN1|SEN2|SEN3|SEN4
Activates the measurement results display of the sensor that is
assigned to the trace.
REF
Selects a reference trace.
HOLD
Freezes the measurement results display of the sensor that is
assigned to the trace.
*RST:
OFF
Example:
:TRACe2:FREQ:SWEep:SRC SEN2
Selects sensor2 for display in trace 2.
:TRACe1:FREQ:SWEep:SRC REF
Assigns the reference trace to trace 1.
:TRACe1:FREQ:SWEep:SRC ON
Activates trace 1 i.e. the reference trace.
Manual operation: See "Indication - Trace Power Analysis" on page 251
:TRACe<ch>:POW:SWEep:SRC <PowSource>
Determines the trace source of a trace for display in power measurement mode.
Parameters:
<PowSource> OFF | SEN1 | SEN2 | SEN3 | SEN4 | HOLD | REF | ON
ON|OFF
Activates ofr deactivates the display of a trace.
SEN1|SEN2|SEN3|SEN4
Activates the measurement results display of the sensor that is
assigned to the trace.
REF
Selects a reference trace.
HOLD
Freezes the measurement results display of the sensor that is
assigned to the trace.
*RST:
OFF
Power sensor measurement subsystems