Remote control commands
R&S
®
SMA100B
537User Manual 1178.3834.02 ─ 09
Example:
:TRACe1:POW:SWEep:SRC SEN1
Selects the connector assigned to the first port for display in
trace 1.
:TRACe2:POW:SWEep:SRC HOLD
Freezes the power sweep measurement results in trace 1.
Manual operation: See "Indication - Trace Power Analysis" on page 251
:TRACe<ch>:TIME:SWEep:SRC <TimeSource>
Determines the trace source of a trace for display in time measurement mode.
Parameters:
<TimeSource> OFF | SEN1 | SEN2 | SEN3 | SEN4 | HOLD | REF | ON
ON|OFF
Activates ofr deactivates the display of a trace.
SEN1|SEN2|SEN3|SEN4
Activates the measurement results display of the sensor that is
assigned to the trace.
REF
Selects a reference trace.
HOLD
Freezes the measurement results display of the sensor that is
assigned to the trace.
*RST:
OFF
Example:
:TRACe2:TIME:SWEep:SRC SEN2
Selects sensor2 for display in trace 2.
:TRACe1:TIME:SWEep:SRC REF
Assigns the reference trace to trace 1.
:TRACe1:TIME:SWEep:SRC ON
Activates trace 1 i.e. the reference trace.
Manual operation: See "Indication - Trace Power Analysis" on page 251
:TRACe<ch>[:POWer]:SWEep:COLor <Color>
Defines the color of a trace.
Parameters:
<Color> INVers | GRAY | YELLow | BLUE | GREen | RED | MAGenta
*RST: trace 1 = YELLow, trace 2 = GREen, trace 3 =
RED.
Example:
TRAC2:SWE:COL GRAY
Trace2 is indicated in gray color.
Manual operation: See "Color - Trace Power Analysis" on page 251
Power sensor measurement subsystems