Remote control commands
R&S
®
SMA100B
619User Manual 1178.3834.02 ─ 09
Example: See Example "Configure and generate an ILS localizer signal"
on page 595.
Manual operation: See "Left Frequency" on page 155
[:SOURce<hw>]:ILS:LOCalizer:MODE <Mode>
Sets the operating mode for the ILS localizer modulation signal.
Parameters:
<Mode> NORM | LLOBe | RLOBe
NORM
ILS localizer modulation is active.
LLOBe
Amplitude modulation of the output signal with the left lobe
(90Hz) signal component of the ILS localizer signal is active.
RLOBe
Amplitude modulation of the output signal with the right lobe
(150Hz) signal component of the ILS localizer signal is active.
*RST: NORM
Example: See Example "Configure and generate an ILS localizer signal"
on page 595.
Manual operation: See "Operating Mode" on page 154
[:SOURce<hw>]:ILS:LOCalizer:PHASe <Phase>
Sets the phase between the modulation signals of the left and right antenna lobe of the
ILS localizer signal.
The zero crossing of the right lobe (150Hz) signal serves as a reference. The angle
refers to the period of the signal of the right antenna lobe.
Parameters:
<Phase> float
Range: -60 to 120
Increment: 0.01
*RST: 0
Example: See Example "Configure and generate an ILS localizer signal"
on page 595.
Manual operation: See "Left/Right Phase" on page 155
[:SOURce<hw>]:ILS:LOCalizer:PRESet
Sets the parameters of the ILS localizer component to their default values (*RST val-
ues specified for the commands).
For other ILS preset commands, see [:SOURce<hw>]:ILS:PRESet on page 598.
SOURce subsystem