Remote control commands
R&S
®
SMA100B
684User Manual 1178.3834.02 ─ 09
AUTO
Selects the optimum sensitivity automatically.
FIXed
Fixes the internal level detector.
*RST: AUTO
Example:
POW:ALC:DSEN FIX
Manual operation: See "Detector Sensitivity" on page 219
[:SOURce<hw>]:POWer:ALC:EDETector:FACTor <DetectorFact>
Sets the attenuation value of the RF coupler.
Parameters:
<DetectorFact> float
Range: -200 to 200
Increment: 0.01
*RST: 16
Example: See Example "Configure the settings for external ALC"
on page 682
Options: R&S SMAB-K726
Manual operation: See "Detector Coupling Factor" on page 219
[:SOURce<hw>]:POWer:ALC:EDETector:LEVel <ReqGenLev>
Sets the maximum power level at the RF output required for compensating the external
ALC coupler and cable losses.
Parameters:
<ReqGenLev> float
Range: -145 to 40
Increment: 0.01
*RST: 20
Example: See Example "Configure the settings for external ALC"
on page 682
Options: R&S SMAB-K726
Manual operation: See "Required Generator Level" on page 219
[:SOURce<hw>]:POWer:ALC:SONCe
Activates level control for correction purposes temporarily.
Example:
POW:ALC OFF
Deactivates automatic level control at the RF output.
POW:ALC:SONC
Executes level control (once).
SOURce subsystem