Remote control commands
R&S
®
SMA100B
682User Manual 1178.3834.02 ─ 09
Example: Configure the settings for external ALC
The example shows a command sequence to set the parameters of the external ALC
mode. It is assumed that you have set up the measurement with the external coupler,
detector diode and the R&S SMA100B as described in Chapter 8.2.2, "How to set up
an external ALC", on page 220.
// Reset the instrument to start from an initial state
*RST; *CLS
// Select the external ALC mode
// configure the detector coupling factor
// and required RF generator level
SOURce:POWer:ALC[:STATe] EALC
SOURce:POWer:ALC:EDETector:FACTor 16
SOURce:POWer:ALC:EDETector:LEVel 20
// Execute the instrument detector offset compensation.
SOURce:POWer:ALC:SONCe
// This function recalculates and compensates internally detected offset voltages
// Activate RF Output1.
Output1:STATe 1
[:SOURce<hw>]:POWer:ALC:MODE?..............................................................................683
[:SOURce<hw>]:POWer:ALC[:STATe].............................................................................. 683
[:SOURce<hw>]:POWer:ALC:DSENsitivity....................................................................... 683
[:SOURce<hw>]:POWer:ALC:EDETector:FACTor.............................................................. 684
[:SOURce<hw>]:POWer:ALC:EDETector:LEVel................................................................ 684
[:SOURce<hw>]:POWer:ALC:SONCe.............................................................................. 684
[:SOURce<hw>]:POWer:ATTenuation:PATTenuator........................................................... 685
[:SOURce<hw>]:POWer:ATTenuation:RFOFf:MODE......................................................... 685
[:SOURce<hw>]:POWer:EMF:STATe................................................................................685
[:SOURce<hw>]:POWer:LBEHaviour............................................................................... 686
[:SOURce<hw>]:POWer:LIMit[:AMPLitude].......................................................................686
[:SOURce<hw>]:POWer:MANual.....................................................................................686
[:SOURce<hw>]:POWer:MODE.......................................................................................687
[:SOURce<hw>]:POWer:POWer......................................................................................687
[:SOURce<hw>]:POWer:STARt....................................................................................... 688
[:SOURce<hw>]:POWer:STOP........................................................................................688
[:SOURce<hw>]:POWer:STEP:MODE............................................................................. 688
[:SOURce<hw>]:POWer:STEP[:INCRement].................................................................... 688
[:SOURce<hw>]:POWer[:LEVel][:IMMediate]:OFFSet........................................................ 689
[:SOURce<hw>]:POWer[:LEVel][:IMMediate]:RCL............................................................. 689
[:SOURce<hw>]:POWer[:LEVel][:IMMediate][:AMPLitude]..................................................690
[:SOURce<hw>]:POWer:RANGe:LOWer?........................................................................ 690
[:SOURce<hw>]:POWer:RANGe:UPPer?.........................................................................690
[:SOURce<hw>]:POWer:RANGe:MIN?.............................................................................691
[:SOURce<hw>]:POWer:RANGe:MAX?............................................................................691
[:SOURce]:POWer:WIGNore...........................................................................................691
SOURce subsystem