12 Injection Port
12.5 Split/Splitless Injection System
124
GC-2010 Plus
12.5.1.2 Parameter list
TEMP
Range: 0.0−450.0 °C, Default: 25.0 °C
Set the temperature of the injection port.
Keep the injection port temperature as low as is practical to prevent of a decrease in the
service life of the temperature sensor. To instantly vaporize the injected sample, set the
injection port temperature which is suitable for the sample. Due to the minute quantity of
sample injected, the sample vaporizes at temperatures lower than the boiling point of
sample elements. To set the injection port maximum temperature limit, refer to “16.6.4
Setting the maximum temperature limits” in “16.6 GC Configuration”.
12.5.1.3 PF menu
PF menu Description
Reference
section
Print Prints the injection port temperature on Chromatopac.
――
Next Inj
Toggles among the setup screeens if two or more injection ports are
installed in this system.
The [INJ] key can also be used to switch screens.
――