13 Detector
13.3 Thermal conductivity detector (TCD)
168
GC-2010 Plus
13.3.3.5 Detector gas program parameters
FLOW RATE
Range: 0.0−20.0 ml/min, Default: 8.0 ml/min
Set the default flow rate and the final flow rate for each stage of the detector gas flow
rate program.
TIME
Range: 0.00−9999.99 min, Default: 0.00 min
Set the retention time for the initial flow rate and the final flow rate for each ramp of the
detector gas flow rate program.
RATE
Range: END/-400.0−400.0 ml/min
2
, Default: END
Set the program rate of the detector gas flow rate program.
If you set the rate to “0”, “END” appears and the program ends at the previous ramp.
Move the cursor to “END” and set any numeric value other than “0” to enter the
pressure and the time for that ramp.
13.3.3.6 PF menu
PF menu Description
Reference
section
Del Line Deletes a line at the current cursor position.
――
Ins Line Inserts a line in line at the current cursor position.
――