13 Detector
13.3 Thermal conductivity detector (TCD)
168
GC-2010 Plus
13.3.3.5 Detector gas program parameters
FLOW RATE
Range: 0.020.0 ml/min, Default: 8.0 ml/min
Set the default flow rate and the final flow rate for each stage of the detector gas flow
rate program.
TIME
Range: 0.009999.99 min, Default: 0.00 min
Set the retention time for the initial flow rate and the final flow rate for each ramp of the
detector gas flow rate program.
RATE
Range: END/-400.0400.0 ml/min
2
, Default: END
Set the program rate of the detector gas flow rate program.
If you set the rate to “0”, “END” appears and the program ends at the previous ramp.
Move the cursor to “END” and set any numeric value other than “0” to enter the
pressure and the time for that ramp.
13.3.3.6 PF menu
PF menu Description
Reference
section
Del Line Deletes a line at the current cursor position.
――
Ins Line Inserts a line in line at the current cursor position.
――

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu GC-2010 Plus and is the answer not in the manual?

Shimadzu GC-2010 Plus Specifications

General IconGeneral
TypeGas Chromatograph
ModelGC-2010 Plus
ManufacturerShimadzu
Maximum Temperature Ramp Rate120 °C/min
Detector OptionsFID, TCD, ECD, FPD, NPD, MS
Electronic Flow Control (EFC)Yes
Data Sampling RateUp to 250 Hz
Column Oven Temperature RangeAmbient +5°C to 450°C
Maximum Oven Heating Rate120 °C/min
Injection PortSplit/Splitless, PTV
DisplayLCD

Related product manuals