16 Special Functions
16.6 GC Configuration
245
GC-2010 Plus
AIR
Selection: Yes/No, Default: Yes
Indicate which detector APC zones should be included in the check.
The displayed items depend on the detector type.
AUX APC
APC1APC18 PRESS/FLOW
Selection: Yes/No, Default: Yes
This item is available only when optional APC units are installed.
DETECTOR
CONTROLLER
Selection: Yes/No, Default: Yes
This item is available only when the FID is installed.
DRIFT
Selection: Yes/No, Default: No
This item compares the baseline drift to the drift limit for 10 minutes.
Once the GC becomes ready, the GC re-evaluates of the ready state 10 minutes later.
WAIT SIGNAL
Selection: Yes/No, Default: Yes
The wait signal applies to all detectors configured in analytical lines.
NOTE User-specified names are used in this screen.
16.6.6.3 PF menu
PF menu Description
Reference
section
Drift Set the baseline drift limit.
――

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu GC-2010 Plus and is the answer not in the manual?

Shimadzu GC-2010 Plus Specifications

General IconGeneral
TypeGas Chromatograph
ModelGC-2010 Plus
ManufacturerShimadzu
Maximum Temperature Ramp Rate120 °C/min
Detector OptionsFID, TCD, ECD, FPD, NPD, MS
Electronic Flow Control (EFC)Yes
Data Sampling RateUp to 250 Hz
Column Oven Temperature RangeAmbient +5°C to 450°C
Maximum Oven Heating Rate120 °C/min
Injection PortSplit/Splitless, PTV
DisplayLCD

Related product manuals