16 Special Functions
16.6 GC Configuration
245
GC-2010 Plus
AIR
Selection: Yes/No, Default: Yes
Indicate which detector APC zones should be included in the check.
The displayed items depend on the detector type.
AUX APC
APC1APC18 PRESS/FLOW
Selection: Yes/No, Default: Yes
This item is available only when optional APC units are installed.
DETECTOR
CONTROLLER
Selection: Yes/No, Default: Yes
This item is available only when the FID is installed.
DRIFT
Selection: Yes/No, Default: No
This item compares the baseline drift to the drift limit for 10 minutes.
Once the GC becomes ready, the GC re-evaluates of the ready state 10 minutes later.
WAIT SIGNAL
Selection: Yes/No, Default: Yes
The wait signal applies to all detectors configured in analytical lines.
NOTE User-specified names are used in this screen.
16.6.6.3 PF menu
PF menu Description
Reference
section
Drift Set the baseline drift limit.
――