17 Printing
17.2 AOC commands
278
GC-2010 Plus
1): Without autosampler, short rack: 1−6
Without autosampler, long rack: 1−12 (1−15 allowable with an additional rack)
With autosampler: 1−150 (maximum; for 1.5 mL vial)
1−96 (maximum; for 4 mL vial)
When an autosampler carousel, the allowable vial No. range depend on the vial rack type and the number of
racks. A value of 0 (default) indicates no sample number to be set.
NOTE Command names marked with “ * ” cannot be set on the GC screen. Use the keys on the AOC-20i.
SAMU Use of the sampler Not use: 0, Use: 1 Not use: 0
SUBU Use of the sub AOC Not use: 0, Use: 1 Not use: 0
BARC Bar Code Reader Not use: 0, Use: 1 Not use: 0
SPMD Distribution of sample for dual AOC 0−80
PAR1 Use of Same Param No: 0, Yes: 1 No: 0
*GLPM Setting of validation mode 0, 1 0
*GRPT Number of times of GLP sample discharge 1−99 50
*GVOL GLP sample discharge volume 1−80 20
*GPMP
Number of times of pumping after second
GLP
0−51
*ATSP Automatic stop function Off: 0, On: 1 On: 1
*ARSG Ready signal polarity Open: 0, Close: 1 Open: 0
TLET Rack Short: 0, Long: 1 Short: 0
TSEL Rack position while sampler is used 0−21
Command
name
Item Range Default