17 Printing
17.3 Program Parameters
280
GC-2010 Plus
Q Time program STOP and repetitions
When a time program STOP value is specified, it is printed with the four digits described
below.
1234
・ Digit 1] value
0: Continuous
When a program finishes, the next program automatically starts without waiting
for the start command.
1: Pause at each run (AOC)
When a program finishes, the GC waits for the start command.
・ Digit 2] and 3] valves
Number of times of run (00−99)
“00” indicates 1 repetition.
When both digits 2] and 3] are set to “0”, nothing is printed but the number of
repetition times is set to “1”.
・ Digit 4] value
File No. to switch to after repetitious run is finished (0−9)
In order to stop the program, input the current file No.
Q Pre-Run program (STRT)
Pre-Run program values are printed with numbers 0−4, described below.
0: Auto
1: Manual (Stop program)
2: GC start
3: AOC/HSS start
4: Clean up