17 Printing
17.4 Event No.
283
GC-2010 Plus
156 Turns On DET #4 makeup gas. Turns Off DET #4 makeup gas.
157 Turns On DET #4 H2. Turns Off DET #4 H2.
158 Turns On DET #4 Air. Turns Off DET #4 Air.
161
Turns On CAR1 gas saver.
Sets split ratio to value of CAR1 gas saver split
ratio 1.
When GC becomes ready, split ratio returns to
value of analysis parameter.
Turns Off CAR1 gas saver.
Returns split ratio to value of analysis
parameter.
162
Turns On CAR1 gas saver.
Sets split ratio to value of CAR1 gas saver split
ratio 2.
When GC becomes ready, split ratio returns to
value of analysis parameter.
Turns Off CAR1 gas saver.
Returns split ratio to value of analysis
parameter.
163
Turns On CAR2 gas saver.
Sets split ratio to value of CAR2 gas saver split
ratio 1.
When GC becomes ready, split ratio returns to
value of analysis parameter.
Turns Off CAR2 gas saver.
Returns split ratio to value of analysis
parameter.
164
Turns On CAR1 gas saver.
Sets split ratio to value of CAR2 gas saver split
ratio 2.
When GC becomes ready, split ratio returns to
value of analysis parameter.
Turns Off CAR2 gas saver.
Returns split ratio to value of analysis
parameter.
171
Turns On CAR1/CAR2/CAR3 high pressure
injection mode.
Turns Off CAR1/CAR2/CAR3 high pressure
injection mode.
Returns column input pressure to value of
analysis parameter.
181 Turns On APC1 flow controller. Turns Off APC1 flow controller.
182 Turns On APC2 flow controller. Turns Off APC2 flow controller.
183 Turns On APC3 flow controller. Turns Off APC3 flow controller.
184 Turns On APC4 flow controller. Turns Off APC4 flow controller.
185 Turns On APC5 flow controller. Turns Off APC5 flow controller.
186 Turns On APC6 flow controller. Turns Off APC6 flow controller.
187 Turns On APC7 flow controller. Turns Off APC7 flow controller.
188 Turns On APC8 flow controller. Turns Off APC8 flow controller.
189 Turns On APC9 flow controller. Turns Off APC9 flow controller.
190 Turns On APC10 flow controller. Turns Off APC10 flow controller.
191 Turns On APC11 flow controller. Turns Off APC11 flow controller.
192 Turns On APC12 flow controller. Turns Off APC12 flow controller.
193 Turns On APC13 flow controller. Turns Off APC13 flow controller.
194 Turns On APC14 flow controller. Turns Off APC14 flow controller.
195 Turns On APC15 flow controller. Turns Off APC15 flow controller.
196 Turns On APC16 flow controller. Turns Off APC16 flow controller.
197 Turns On APC17 flow controller. Turns Off APC17 flow controller.
198 Turns On APC18 flow controller. Turns Off APC18 flow controller.
No. Meaning of “Event xx” (Example: Event 1) Meaning of “Event-xx” (Example: Event -1)