19 Troubleshooting
19.2 Error Messages
319
GC-2010 Plus
File operation errors
These messages appear when an incorrect file operation was attempted.
Press another key, and continue operation.
Overflow of calculated pressure value
The carrier gas pressure calculated from the linear velocity, the flow rate or the split ratio.
You have input is outside the set range. Change the conditions so that the pressure is within
the set range, and enter the new value.
Overflow of calculated flow rate value
The carrier gas total flow rate calculated from the split ratio or the pressure you have input is
outside the set range. Change the conditions, and enter a new value.
Overflow of calculated APC pressure value
The pressure calculated from the flow rate you have input is outside the set range. Change
the conditions, and enter a new value.
Overflow of calculated flow rate value in programs
The carrier gas total flow rate calculated from the split ratio program you have input is out-
side the set range.
Code Message CS Note
5002 Invalid file no. No
5003 This file is now used No
5004 File initialize failure No
5005 File can
’t be copied No
Code Message CS Note
5006-5008
5070-5072
CARx calc. prss out of range No x=1-3
Code Message CS Note
5009-5011 CARx calc. flow out of range No x=1-3
Code Message CS Note
5012, 5015
5018, 5021
5045-5048
MUPn calc. prss out of range No
n=1-4
make up gas
5013, 5016
5019, 5022
HGNn calc. prss out of range No
n=1-4
Hydrogen
5014, 5017
5020, 5023
AIRn calc. prss out of range No
n=1-4
Air
5024-5041
5049-5066
APCy calc. prss out of range No
y=1-18
AUX APC
5042-5044
5067-5069
PURx calc. prss out of range No
x=1-3
septum purge
Code Message CS Note
5073, 5074
5077, 5078
CAR m calc.flow out of range No m=1-2
5075, 5076 CAR m calc.ratio out of range No m=1-2