EasyManuals Logo

Shimadzu GC-2010 Plus User Manual

Shimadzu GC-2010 Plus
352 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #63 background imageLoading...
Page #63 background image
3 AFC, APC
3.5 AFC and APC Control Ranges
43
GC-2010 Plus
The allowable pure flow rate setting depends on the column inlet pressure. Figure 3.5.3
shows the range of purge flow rates permitted.
Fig. 3.5.3  
Allowable APC settings are shown below, for a supply pressure of 500 kPa. (When the
supply pressure is less than 500 kPa, the available maximum flow rate decreases
proportionally.) When the set value is more than the set range, displayed values may not be
accurate, or APC cannot control the flow.
Air (detector = FID) : 0 to 1,000 ml/min
H
2 (detector = FID) : 0 to 100 ml/min
Makeup gas (detector = FID) : 0 to 100 ml/min
Makeup gas (detector = TCD) : 0 to 20 ml/min
0
1
2
3
4
5
6
02468101214161820
0
0
50
100
150
200
250
300
100 200 300 400 500
Column inlet pressure (kPa)
Column inlet pressure (kPa)
Setting range
Setting range
Purge flow rate (ml/min)Purge flow rate (ml/min)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu GC-2010 Plus and is the answer not in the manual?

Shimadzu GC-2010 Plus Specifications

General IconGeneral
TypeGas Chromatograph
ModelGC-2010 Plus
ManufacturerShimadzu
Maximum Temperature Ramp Rate120 °C/min
Detector OptionsFID, TCD, ECD, FPD, NPD, MS
Electronic Flow Control (EFC)Yes
Data Sampling RateUp to 250 Hz
Column Oven Temperature RangeAmbient +5°C to 450°C
Maximum Oven Heating Rate120 °C/min
Injection PortSplit/Splitless, PTV
DisplayLCD

Related product manuals