7 Starting and Stopping the GC [SYSTEM]
7.2 Specifying Clean Up Parameters
76
GC-2010 Plus
Q Clean up column oven temperature program
(The clean up temperature program consists of a single program ramp.)
CLEAN UP RATE
Range: END/-250.00250.00 °C/min, Default: END
Sets the rate of column temperature increase for the clean up program.
Refer to “1.2.1 Column oven”
.
CLEAN UP TEMP
Range: 0.0450.0 °C, Default: 25.0 °C
Sets the final temperature for the column oven temperature clean up program.
Do not exceed the maximum column temperature.
CLEAN UP TIME
Range: 0.009999.99 min, Default: 0.00 min
Sets the final temperature hold time for the clean up program.
Q Clean up column Inlet pressure program
(The clean up pressure program consits of a single program ramp.)
CLEAN UP RATE
Range: END/-400.00400.00 kPa/min, Default: END
Sets the rate of column inlet pressure for the clean up program.
The rate of pressure depends on the total flow rate.
CLEAN UP PRESS
Range: 0.0970.0 kPa (Refer to Fig. 3.5.1.) , Default: 0.0 kPa
Sets the final pressure for the column inlet pressure clean up program.
CLEAN UP TIME
Range: 0.009999.99 min, Default: 0.00 min
Sets the final pressure hold time for the clean up program.
Q Clean up total flow rate program
(The clean up total flow rate program consists of a single program ramp.)
CLEAN UP RATE
Range: END/-400.00400.00 ml/min
2
, Default: END
Sets the rate of total flow increase for the clean up program.
CLEAN UP FLOW RATE
Range: 0.01200.0 ml/min (Refer to Fig. 3.5.1.), Default: 50.0 ml/min
Set the final flow rate for the total flow rate clean up program.
CLEAN UP TIME
Range: 0.009999.99 min, Default: 0.00 min
Sets the final flow rate hold time for the clean up program.