Setup
3.4 Inputs and outputs
FST030
Function Manual, 04/2017, A5E35693328-AA
149
Setting
289 Sound velocity simulated; 290 Flow velocity simulated; 291 Pressure simulated; 292
Kin. viscosity simulated; 293 Aux. temperature simulated; 294 Sensor temperature simu-
lated; 296 CH5 simulated; 297 CH6 simulated; 288 Std. volume flow simulated; 299
Standardizing factor simulated; 300 Std. Kin. viscosity simulated; 301 Std. density simulat-
ed; 302 LiquIdent simulated; 303 API gravity simulated; 304 Std. API gravity simulated;
305 Specific gravity simulated; 306 Std. spec. grav. simulated; 307 Rate of change simu-
lated
Default
None
Related
Path
Setup → Inputs and outputs → Channel 3 - input/output → Status output → Simulation
alarms (2)
3.4.5.8.15 Alarm class
Description
Selection of alarm classes which impact the status output. Each diagnostic event is allocated
to an alarm class. Applicable if Status mode = Alarm class and Status icons = Standard.
Bit 0: Process value alarm (PA)
Bit 1: Process value warning (PW)
Bit 2: Maintenance alarm (MA)
Bit 3: Maintenance warning (MW)
Bit 4: Maintenance required (MR)
Bit 5: Function check (FC)
Setting
Maintenance alarm; Function check; Process value alarm; Process value warning; Mainte-
nance demanded; Maintenance required; None
Default
Related
Path
Setup → Inputs and outputs → Channel 3 - input/output → Status output → Alarm class
3.4.5.8.16 NAMUR status signal
Description
Selection of NAMUR status signals which impact the status output. Each diagnostic event is
allocated to a NAMUR status signal. Application if Status mode = Alarm class and Status
icons = NAMUR
Setting
Failure; Function check; Out of specification; Maintenance required
Default
None