Syntax
AUXIn:PRObe:D
EGAUSS EX ECute
Arguments
EXECUTE starts a probe degauss cycle.
Examples
AUXIN:PROBE:DEGAUSS EXECUTE, degausses the probe attached to the AUX
In input.
AUXIn:PRObe:DEGAUSS:STATE? (Query Only)
This command queries whether the probe attached to the AUX In input has
been degaussed.
Group
Vertical
Syntax
AUXIn:PRObe:DEGAUSS:STATE ?
Examples
AUXIN:PROBE:DEGAUSS:STATE ? might return
AUXIN:PROBE:DEGAUSS:STATE PASSED, indicating that the probe has been
degaussed.
AUXIn:PRObe:FORCEDRange
This command sets the probe a
ttached to the AUX In input to the specied range,
or the command queries the probe range.
Group
Vertical
Syntax
AUXIn:PRObe:FORCEDRange <NR3>
AUXIn:PRObe:FORCEDRange?
Arguments
<NR3> species the probe range.
Examples
If a TCP0030 current probe is attached the AUX IN input, AUXIN :PROBE:
FORCEDRANGE 5.0
sets the probe to its 5 Ampere range.
AUXIN:PROBE:FORCEDRANGE might return :AUXIN:PROBE:FORCEDRANGE
10.0000E+36
, indicating that the probe range is set to 10.0000E+36 dB.
2-90 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual

Table of Contents

Other manuals for Tektronix DSA70000/B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DSA70000/B Series and is the answer not in the manual?

Tektronix DSA70000/B Series Specifications

General IconGeneral
BrandTektronix
ModelDSA70000/B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals