Commands Listed in Alphabetical Order
CURSor:XY:YD
ELta? (Query Only)
This query-only command returns the XY cursor Y delta value in rectangular
coordinates.
NOTE. This q
uery is only valid when the CURSor:XY:READOUT is set to
RECTangular. If the readout is set to other than rectangular coordinates, the
IEEE representation of NAN (Not a Number) is returned: 99.1000E+36.
Group
Cursor
Syntax
CURSor:XY:YDELta?
Examples
CURSOR:XY:YDELTA? might return :CURSOR: XY:YDELTA 40.0000E-6,
indicating that the DY value is 40 μV.
CURVe
This command transfers waveform data to and from the instrument. Each
waveform that is transferred has an associated waveform preamble that contains
information such as data format and scale.
The CURVe? query transfers data from the instrument. The data source is
specified by the DATa:SOUrce command. The first and last data points are
specified by the DATa:STARt and DATa:STOP commands.
The CURVe command transfers waveform data to the instrument. The data
is stored in the reference memory location specified by DATa:DESTination,
starting with the data point specified by DATa:STARt. Only one waveform can
be transferred at a time. The waveform will only be displayed if the reference
is displayed.
Group
Waveform Transfer
Syntax
CURVe { <Block>|<asc curve>}
CURVe?
Related Commands
DATa:DESTination, DATa:SOUrce, DATa:STARt, DATa:STOP,
SAVe:WAVEform, SAVe:WAVEform:FILEFormat, WFMInpre?,
WFMInpre:BYT_Nr, WFMOutpre?, WFMInpre:NR_FR, WFMOutpre:NR_FR?
2-150 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual