Commands Listed in Alphabetical Order
TRIGGER:A:LOG
IC:PATTERN:INPUT:CH4?
might return
:TRIGGER:A:LOGIC:PATTERN:I NPUT:CH4 HIGH, indicating that
the logic input for Channel 4 is logic high.
TRIGger:{A|B}:LOGIc:PATtern:WHEn
This comma
nd sets or queries the condition for generating an A or B logic pattern
trigger with respect to the dened input pattern. This command is equivalent to
selecting A or B Event (Main) Trigger Setup from the Trig menu, selecting Pattern
for Trigger Type, and choosing a trigger condition from the Pattern drop-down
list, which is located in the Trigger When group box.
Group
Trigger
Syntax
TRIGger:{A|B}:LOGIc:PATter n:WHEn
{TRUe|FALSe|LESSThan|MORET han}
TRIGger:{A|B}:LOGIc:PATter n:WHEn?
Related Commands
TRIGger:{A|B}:LOGIc:FUNCtion, TRIGger:{A|B}:LOGIc:INPut?, TRIGger:
{A|B}:LOGIc:THReshold?, TRIGger:{A|B}:LOGIc:PATtern:WHEn:LESSLimit,
TRIGger:{A|B}:LOGIc:PATtern:WHEn:MORELimit
Arguments
TRUe sets the instrument to trigger when the pattern becomes true.
FALSe sets the instrument to trigger when the pattern becomes false.
LESSThan sets the instrument to trigger if the specic pattern is true less than the
time set by the TRIGger:{A|B}:LOGIc:PATtern:WHEn:LESSLimit command.
MOREThan argument sets the instrument to trigger if the specic pattern is true
longer than the specied time set by the TRIGger:{A|B}:LOGIc:PATtern:WHEn:
MORELimit command.
Examples
TRIGGER:A:LOGIC:PATTERN:WH EN FALSE species to trigger the A logic
pattern when the pattern becomes false.
TRIGGER:A:LOGIC:PATTERN:WH EN? might return
:TRIGGER:A:LOGIC:PATTERN:W HEN TRUE, indicating that the A logic pattern
will trigger when the pattern becomes true.
TRIGger:{A|B}:LOGIc:PATtern:WHEn:LESSLimit
This command sets or queries the maximum time that the selected pattern may be
true and still generate an A or B logic pattern trigger. This command is equivalent
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-507

Table of Contents

Other manuals for Tektronix DSA70000/B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DSA70000/B Series and is the answer not in the manual?

Tektronix DSA70000/B Series Specifications

General IconGeneral
BrandTektronix
ModelDSA70000/B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals