Commands Listed in Alphabetical Order
Examples
TRIGGER:A:LOG
IC:SETHOLD:DATA?
might return
:TRIGGER:A:LOGIC:SETHOLD:D ATA:THRESHOLD 1.4000;SOURCE CH1,
indicating the current trigger data settings.
TRIGger:{A|
B}:LOGIc:SETHold:DATa:SOUrce
This command sets or queries the data source for the setup and hold trigger. This
command is equivalent to selecting Setup/Hold Setup from the Trig menu and
choosing t
he desired channel from the Data Source drop-down list.
Group
Trigger
Syntax
TRIGger:{A|B}:LOGIc:SETHol d:DATa:SOUrce CH<x>
TRIGger:{A|B}:LOGIc:SETHol d:DATa:SOUrce?
Related Commands
TRIGger:{A|B}:LOGIc:SETHold:DATa:THReshold
Arguments
CH<x> specifies the input channel, which ranges from 1 through 4.
Examples
TRIGGER:A:LOGIC:SETHOLD:DA TA:SOURCE CH1 sets channel 1 as the clock
source for the setup and hold trigger.
TRIGGER:A:LOGIC:SETHOLD:DA TA:SOURCE? might return
:TRIGGER:A:LOGIC:SETHOLD:D ATA:SOURCE CH2 indicating that Channel 2
is the current clock source for the setup and hold trigger.
TRIGger:{A|B}:LOGIc:SETHold:DATa:THReshold
This command sets or queries the data voltage threshold for setup and hold trigger.
This command is equivalent to selecting Setup/Hold Setup from the Trig menu
and then setting the desired Data Level.
Group
Trigger
Syntax
TRIGger:{A|B}:LOGIc:SETHol d:DATa:THReshold {ECL|TTL |<NR3>}
TRIGger:{A|B}:LOGIc:SETHol d:DATa:THReshold?
Related Commands
TRIGger:{A|B}:LOGIc:SETHold:DATa:SOUrce
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-513