Commands Listed in Alphabetical Order
LIMit:LOCk
This command sets or queries whether vertical scaling and positioning affect both
source and template for template comparison pairs.
NOTE. The co
mmand works only if the LT Limit Test option is enabled.
Group
Limit Test
Syntax
LIMit:LOCk {OFF|ON|<NR1>}
LIMit:LOCk?
Arguments
OFF disables the lock feature.
ON enables the lock feature.
<NR1> = 0 disables the lock feature; any other value enables the lock feature.
For queries, a 0 is returned if the lock feature is off; a 1 is returned if the lock
feature is on.
Examples
LIMIT:LOCK ON enables the lock feature so that vertical scaling and positioning
affect both the source and the template for template comparison pairs.
LIMIT:LOCK? might return :LIMIT:LOCK 0, indicating that the lock feature is
disabled.
LIMit:LOG
This command sets or queries whether a log le is saved when the source
waveform data exceeds the test limits.
NOTE. The command works only if the LT Limit Test option is enabled.
Group
Limit Test
Syntax
LIMit:LOG {OFF|ON|<NR1>}
LIMit:LOG?
Arguments
OFF disables saving a log le when the s ource waveform exceeds the test limits.
ON enables saving a log le when the source waveform exceeds the test limits.
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-271

Table of Contents

Other manuals for Tektronix DSA70000/B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DSA70000/B Series and is the answer not in the manual?

Tektronix DSA70000/B Series Specifications

General IconGeneral
BrandTektronix
ModelDSA70000/B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals