Commands Listed in Alphabetical Order
AUXIn:PRObe:
INPUTMode
This command sets or returns one of the probe’s four input modes on the AUX
In input.
Group
Vertical
Syntax
AUXIn:PRObe:INPUTMode
{makeEnumList(signalPathM gr_probeTipMode}
AUXIn:PRObe:INPUTMode?
Arguments
DEFault sets the probe to the default mode.
DIFFerential sets the probe to route differential signals to the host.
COMmonmode sets the probe to route common-mode signals to the host.
A sets the probe to route single-ended A signals to the host.
B sets the probe to route single-ended B signals to the host.
Examples
AUXIN:PROBE:INPUTMODE A sets the probe to route single-ended A signals to
the host.
AUXIN:PROBE:INPUTMODE? might return :AUXIN:PROBE:INPUTMODE
DEFAULT
, indicating that the probe input mode on the AUXIN input is set to
DEFAULT.
AUXIn:PRObe:INPUTMode:AOFFSet
This command sets or returns the requested A mode offset control of the probe
that is attached to the AUX In input.
Group
Vertical
Syntax
AUXIn:PRObe:INPUTMode:AOF FSet
AUXIn:PRObe:INPUTMode:AOF FSet?
Arguments
NR3 species the A mode offset control.
Examples
AUXIN:PROBE:INPUTMODE:AOF FSET 0.2, sets the A m ode offset control to
0.2 V.
2-92 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual

Table of Contents

Other manuals for Tektronix DSA70000/B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DSA70000/B Series and is the answer not in the manual?

Tektronix DSA70000/B Series Specifications

General IconGeneral
BrandTektronix
ModelDSA70000/B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals