EasyManuals Logo

Tektronix DSA70000/B Series Programmer's Manual

Tektronix DSA70000/B Series
733 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #319 background imageLoading...
Page #319 background image
Commands Listed in Alphabetical Order
<NR1> =0turnso
ff mask lock to waveform and any other value turns on mask
lock to waveform.
Examples
MASK:LOCK ON
turnsonmasklocktowaveform.
MASK:LOCK? might return :MASK:LOCK 1, indicating that the mask lock to
waveform mode is enabled.
MASK:MARgin:PERCent
This command sets or queries the mask ma rgin percentage. If you set a positive
mask margin percentage, the polygon expands a nd the mask test will be more
difcult to pass due to the expanded test area. If you set a negative mask m argin
percentage, the polygon shrinks and the mask test will be easier to pass due to
the
smaller area.
Group
Mask
Syntax
MASK:MARgin:PERCent <NR3>
MASK:MARgin:PERCent?
Arguments
<NR3> values range from -50.0 to +50.0. The default is 5.
Examples
MASK:MARGIN:PERCENT 25 sets the mask margin percentage to 25.
MASK:MARGIN:PERCENT? might return :MASK:MARGIN: PERCENT
1.00000000000E+001,
indicating that the mask margin is set to 10%.
MASK:MARgin:STATE
This command s ets o r returns the state of the mask margins.
Group
Mask
Syntax
MASK:MARgin:STATE {OFF|ON|<NR1>}
MASK:MARgin:STATE?
Argumen
ts
ON turns on mask margins.
OFF turns off mask margins. The currently displayed margined mask is erased and
the original mask is displayed.
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-307

Table of Contents

Other manuals for Tektronix DSA70000/B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DSA70000/B Series and is the answer not in the manual?

Tektronix DSA70000/B Series Specifications

General IconGeneral
BrandTektronix
ModelDSA70000/B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals