EasyManuals Logo

Tektronix DSA70000/B Series Programmer's Manual

Tektronix DSA70000/B Series
733 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #566 background imageLoading...
Page #566 background image
Commands Listed in Alphabetical Order
TRIGger:A:CA
N:DATa:VALue
CAN option only: This command sets or quires the binary data string used for
CAN Trigger if the trigger condition is ID or IDANDDATA.
Group
Low Speed Serial Trigger
Syntax
TRIGger:A:CAN:DATa:VALue <String>
TRIGger:A:CAN:DATa:VALue?
Arguments
String is up to 32 bits specifying the CAN data value.
Examples
TRIGGER:A:CAN:DATA:VALUE “1011” sets the CAN data value to 1011.
TRIGGER:A:CAN:DATA:VALUE? might return
:TRIGGER:A:CAN:DATA:SOURC E 10111, indicating that the CAN data value
is set to 10111.
TRIGger:A:CAN:FORMat
CAN option only: This command sets or queries the CAN data format.
Group
Low Speed Serial Trigger
Syntax
TRIGger:A:CAN:FORMat {BINary|HEX}
TRIGger:A:CAN:FORMat?
Arguments
BINary species binary as the CAN data format.
HEX spec ies hexadecimal as the CAN data format.
Examples
TRIGGER:A:CAN:FORMAT BINARY sets the CAN trigger data format to bin
ary.
TRIGGER:A:CAN:FORMAT? might return :TRIGGER:A:CAN:DATA:FORMAT
HEX
, indicating that the CAN data format is set to hexadecimal.
TRIGger:A:CAN:FRAMEtype
CAN option only: This command sets or queries the CAN trigger frame type.
2-554 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual

Table of Contents

Other manuals for Tektronix DSA70000/B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DSA70000/B Series and is the answer not in the manual?

Tektronix DSA70000/B Series Specifications

General IconGeneral
BrandTektronix
ModelDSA70000/B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals