Commands Listed in Alphabetical Order
Syntax
TRIGger:B:BY {
EVENTS|TIMe}
TRIGger:B:BY?
Related Commands
TRIGger:B:EVENTS:COUNt, TRIGger:B:TIMe, HORizontal[:MAIn]:DELay:
MODe, HORizontal[:MAIn]:DELay:TIMe TRIGger:B:STATE
Arguments
EVENTS sets the B trigger to take place following a set number of trigger
events after the A trigger occurs. The number of events is specied by
TRIGger:B:EVENTS:COUNt.
TIMe sets the B trigger to occur a set time after the A trigger event. The time
period is specied by
TRIGger:B:TIMe.
Examples
TRIGGER:B:BY TIME sets the B trigger to occur at a set time after the A trigger
event.
TRIGGER:B:BY? might return :TRIGGER:B:BY EVENTS, indicating that the B
trigger takes place following a set number of trigger events a fter the A trigger
occurs.
TRIGger:B:EVENTS? (Query Only)
This query-only command returns the current B trigger events parameter. This
command is equivalent to selecting B Event (Delayed) Trigger Setup from the
Trig menu, selecting the A B Seq tab, choosing Trig on nth event, and viewing
the Trig Event setting.
Group
Trigger
Syntax
TRIGger:B:EVENTS?
Related Commands
TRIGger:B:EVENTS:COUNt, TRIGger:B:STATE
Examples
TRIGGER:B:EVENTS? might return
:TRIGGER:B:EVENTS:COUNT 2, indicating that the two events must occur
before the B trigger occurs.
TRIGger:B:EVENTS:COUNt
This command sets or queries the number of events that must occur before the B
trigger (when TRIG:DELay:BY is set to EVENTS). This command is equivalent
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-609