Commands Listed in Alphabetical Order
Examples
MASK:HIGHLIGH
THITS ON
enables the mask highlighthits function.
MASK:HIGHLIGHTHITS? might return :MASK:HIGHL IGHTHITS 1, indicating
that the mask:
highlighthits function is enabled.
MASK:INVert
This command controls whether the mask is drawn inverted. It has no effect if this
mask cannot be inverted. The default is Off (Positive).
Group
Mask
Syntax
MASK:INVert {OFF|ON|<NR1>}
Arguments
ON Negative
OFF Positive
<NR1> = 0 disables the mask:invert function; any other value enables it.
Examples
MASK:INVERT ON enables the mask invert function.
MASK:INVERT? might return :MASK:INVERT 1, indicating that the mask is
drawn inverted.
MASK:LOCk
This command sets or returns the mask lock to waveform mode. It locks the mask
to the waveform such that changes to the horizontal and/or vertical scale settings
redraws the mask segments in proportion. This feature allows for expanding
horizontal and/or vertical settings to zoom in on waveforms and masks to visually
examine violation areas in more detail.
Group
Mask
Syntax
MASK:LOCk {OFF|ON|<NR1>}
MASK:LOCk?
Arguments
ON turnsonmasklocktowaveform.
OFF turns off mask lock to waveform.
2-306 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual