Commands Listed in Alphabetical Order
Examples
LIMIT:BEEP ON c
auses a beep when any waveform data exceeds the limits
specied in the limit test.
LIMIT:BEEP? m
ight return
:LIMIT:BEEP 0, indicating that the instrument will
not beep when waveform data exceeds the limits specied in the limit test.
LIMit:COMpare (No Query Form)
This command resets the Limit Test comparison template.
NOTE. The command works only if the LT Limit Test option is enabled.
Group
Limit Test
Syntax
LIMit:COMpare {RESet}
Arguments
RESet resets the Limit template to “None.”
Examples
LIMIT:COMPARE RESET sets the Limit template to None.
LIMit:COMpare:CH<x>
This command sets or queries the template against w hich to compare the
waveform acquired from the channel specied by x. The template can be a
waveform saved in any of the reference locations (REF1 through REF4) or none.
NOTE. The command works only if the LT Limit Test option is enabled.
Group
Limit Test
Syntax
LIMit:COMpare:CH<x> {NONe|REF<x>}
LIMit:COMpare:CH<x>?
Arguments
NONe turns off template testing for the channel specied by CH<x>.
CH<x> selects which channel waveform to use as the template against which to
compare the waveforms acquired by the specied channel (CH<x>).
2-266 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual

Table of Contents

Other manuals for Tektronix DSA70000/B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DSA70000/B Series and is the answer not in the manual?

Tektronix DSA70000/B Series Specifications

General IconGeneral
BrandTektronix
ModelDSA70000/B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals