Commands Listed in Alphabetical Order
Group
Search and Mark
Syntax
SEARCH:SEARCH<x>:TRIGger: A:RUNT:THReshold:LOW:REF<x> <NR3>
SEARCH:SEARCH<x>:TRIGger: A:RUNT:THReshold:LOW:REF<x>?
Arguments
<NR3> specifies the LOW threshold in volts.
Examples
SEARCH:SEARCH1:TRIGGER:A: RUNT:THRESHOLD:LOW:REF2 0 .10 sets the
Reference 2 waveform runt trigger LOW threshold for search 1 to 100 mV.
SEARCH:SEARCH1:TRIGGER:A: RUNT:THRESHOLD:LOW:REF1? might
return
:SEARCH:SEARCH1:TRIGGER:A: RUNT:THRESHOLD:LOW:REF1
800.0000E-3
, indicating that the Reference 1 waveform runt trigger LOW
threshold for search 1 is 800 mV.
SEARCH:SEARCH<x>:TRIGger:A:RUNT:WHEn
This command sets or returns the
condition setting for a runt trigger search to
determine where to place a mark. <x> is the search number.
Group
Search and Mark
Syntax
SEARCH:SEARCH<x>:TRIGger: A:RUNT:WHEn {OCCurs|WIDER than}
SEARCH:SEARCH<x>:TRIGger: A:RUNT:WHEn?
Arguments
OCCurs a rgument specifies a trigger event if a runt of any d etectable width occurs.
WIDERthan argument sets the oscilloscope to trigger if a runt pulse is detected
with width wider than the time set by the SEARCH:SEARCH<x>:TRIGger:
A:RUNT:WIDth command.
Examples
SEARCH:SEARCH1:TRIGGER:A: RUNT:WHEN WIDERTHAN sets the oscilloscope
to trigger when a runt pulse is detected with width w ider than the time set by the
SEARCH:SEARCH<x>:TRIGger:A:RUNT:WIDth command.
SEARCH:SEARCH1:TRIGGER:A: RUNT:WHEN? might return
:SEARCH:SEARCH1:TRIGGER:A :RUNT:WHEN OCCURS, indicating
thatamarkisplacedifarunt trigger event occurs.
2-460 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual