Commands Listed in Alphabetical Order
ZOOm:REF<x>:
VERTical:SCAle
This command sets or queries the zoom vertical scale of the specied reference
waveform.
Group
Zoom
Syntax
ZOOm:REF<x>:VERTical:SCAle <NR3>
ZOOm:REF<x>:VERTical:SCAle ?
Arguments
<NR3> sets the horizontal scale factor of the specied Reference 1 waveform to 10.
Examples
ZOOM:REF1:VERTICAL:SCALE 5 sets the vertical scale factor of the specied
Reference 1 waveform to 5.
ZOOM:REF1:VERTICAL:SCALE? might return :ZOOM: REF1:VERTICAL:SCALE
2.0000
, indicating that the vertical scale factor of the specied Reference 1
waveform is set to 2.
ZOOm:SCROLL:DIREction
This command sets o r returns the direction for automatic scrolling of zoomed
waveforms.
Group
Zoom
Syntax
ZOOm:SCRO
LL:DIREction {FWD|FFWD|REV| FREV|STOP}
ZOOm:SCROLL:DIREction?
Arguments
FWD starts AutoScroll. The Zoom Box moves from left to right.
FFWD starts AutoScroll. Increases the scrolling speed in the forward direction.
REV starts AutoScroll. The Zoom Box moves from right to left.
FREV starts AutoScroll. Increases the scrolling speed in the reverse direction.
STOP halts AutoScroll.
Examples
ZOOM:SCROLL:DIRECTION FWD starts AutoScroll. The Zoom Box moves from
left to right.
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-655