Commands Listed in Alphabetical Order
Group
Search and Mark
Syntax
SEARCH:SEARCH<x>:TRIGger: A:LOGIc:THReshold:MATH<x> <NR3>
SEARCH:SEARCH<x>:TRIGger: A:LOGIc:THReshold:MATH<x>?
Arguments
<NR3> species the threshold level in volts.
Examples
SEARC
H:SEARCH1:TRIGGER:A:LOGIC:T HRESHOLD:MATH2 2.
0 sets the
Math 2 waveform logic trigger threshold for search 1 to 2.0 Volts.
SEAR
CH:SEARCH1:TRIGGER:A:LOGIC:T HRESHOLD:MATH1?
might return
:SEARCH:SEARCH1:TRIGGER:A :LOGIC:THRESHOLD:MATH1 1. 4000,
indicating the M ath 1 waveform logic threshold for search 1 is set to 1.4 Volts.
SE
ARCH:SEARCH<x>:TRIGger:A:LOGIc:THReshold:REF<x>
This command sets or queries the reference waveform threshold level for a pattern
or state trigger search to determine where to place a mark.
<x> is the search
n
umber or the reference waveform number.
Group
Search and Mark
Syntax
SEARCH:SEARCH<x>:TRIGger: A:LOGIc:THReshold:REF<x> <NR3>
SEARCH:SEARCH<x>:TRIGger: A:LOGIc:THReshold:REF<x>?
Arguments
<NR3> species the threshold level in volts.
Examples
SEARCH:SEARCH1:TRIGGER:A: LOGIC:THRESHOLD:REF2 2.0 sets the
Reference 2 waveform logic trigger threshold for search 1 to 2.0 Volts.
SEARCH:SEARCH1:TRIGGER:A: LOGIC:THRESHOLD:REF1? might return
:SEARCH:SEARCH1:TRIGGER:A :LOGIC:THRESHOLD:REF1 1.4 000,
indicating the Reference 1 waveform logic trigger threshold for search 1 is set
to 1.4 Volts.
SEARCH:SEARCH<x>:TRIGger:A:PATtern:WHEn
This command sets or queries the pattern trigger pattern test that determines where
toplaceamark.