Commands Listed in Alphabetical Order
TRIGger:{A|B
}:PULse:WINdow:THReshold:BOTh (No Q uery Form)
This command (no query form) sets the upper and lower switching thresholds for
the window trigger. This command is equivalent to selecting Window Setup from
the Trig menu
and then setting the Upper Level and Lower Level voltages.
Group
Trigger
Syntax
TRIGger:{A|B}:PULse:WINdo w:THReshold:BOTh {TTL|ECL}
Arguments
TTL argument sets the upper and the lower threshold to the nominal TTL voltage
levels.
ECL argument sets the upper and the lower threshold to the nominal ECL voltage
levels.
Examples
TRIGGER:A:PULSE:WINDOW:TH RESHOLD:BOTH TTL sets the threshold of the
window trigger to the nominal TTL voltage levels.
TRIGger:{A|B}:PULse:WINdow:THReshold:HIGH
This command sets or queries the upper limit for the pulse window trigger. This
command is equivalent to selecting Window Setup from the Trig menu and setting
the window trigger Upper Level voltage.
Group
Trigger
Syntax
TRIGger:{A|B}:PULse:WINdo w:THReshold:HIGH <NR3>
TRIGger:{A|B}:PULse:WINdo w:THReshold:HIGH?
Arguments
<NR3> specifies the threshold value in volts.
Examples
TRIGGER:A:PULSE:WINDOW:TH RESHOLD:HIGH 2.0, sets the upper limit of the
pulse window trigger to 2 V.
TRIGGER:A:PULSE:WINDOW:TH RESHOLD:HIGH? might return
:TRIGGER:A:PULSE:WINDOW:T HRESHOLD:HIGH 1.1000E+00, indicating that
the upper limit of the window trigger is set to 1.1 V.
2-548 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual