Commands Listed in Alphabetical Order
Examples
FACTORY resets
the instrument to its factory default settings.
FASTAcq? (Query Only)
This query-only command returns the state of Fast Acquisitions. This command is
equivalent t
o pressing the FASTACQ button on the front panel.
Group
Acquisition
Syntax
FASTAcq?
Examples
FASTACQ? might return the following string for the current acquisition parameters:
:FASTACQ:STATE 0.
FASTAc
q:HIACQRATE
This command s ets or queries the state of FastAcq optimization for capturing
the details with a higher sample rate.
Group
Acquisition
Syntax
FASTAcq:HIACQRATE {<NR1>|OFF|ON}
FASTAcq:HIACQRATE?
Arguments
ON sets FastAcq o ptimization for capturing rare events with more acquisitions.
OFF sets FastAcq optimization for capturing details with a higher sample rate.
<NR1> = 0 sets FastAcq optimization for capturing details with a higher sample
rate; any other value sets FastAcq optimization for capturing rare events with
more acquisitions.
Examples
FASTACQ:HIACQRATE ON sets optimizes FastAcq for capturing rare events with
more acquisitions.
FASTACQ:HIACQRATE? might return :FASTACQ:HIACQRATE 0, indicating that
the FastAcq is optimized for capturing details with a higher sample rate.
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-217