Commands Listed in Alphabetical Order
Group
Trigger
Syntax
TRIGger:A:HOLDoff?
Related Commands
TRIGger:A:HOLDoff:ACTUal?, TRIGger:A:HOLDoff:BY, TRIGger:A:
HOLDoff:TIMe
Examples
TRIGGER:A:HOLDOFF? might return :TRIGGER:A:HOLDOFF:TIME
900.0000E-09;BY DEFAULT
, indicating that the A edge trigger holdoff time
(by default) is set to 900 ns.
TRIGger:A:HOLDoff:ACTUal? (Query Only)
This query-only command returns the holdoff time actually used (expressed in
seconds) by the A trigger. This command is equivalent to selecting Holdoff from
the Trig menu and then viewing the current Trig Holdoff value.
Group
Trigger
Syntax
TRIGger:A:HOLDoff:ACTUal?
Related Commands
TRIGger:A:HOLDoff?, TRIGger:A:HOLDoff:BY, TRIGger:A:HOLDoff:TIMe
Examples
TRIGGER:A:HOLDOFF:ACTUAL? migh
treturn
:TRIGGER:A:HOLDOFF:ACTUAL
4.0000E-06
, indicating that the holdoff time is s et to 4 µs.
TRIGger:A:HOLDoff:BY
This command sets or queries the type of holdoff for the A trigger. Holdoff types
are expressed as either user-specified time (TIMe) or by an internally calculated
minimum time value (DEFAult/AUTO). This command is equivalent to selecting
Holdoff from the Trig menu and t
hen s etting the Holdoff type.
Group
Trigger
Syntax
TRIGger:A:HOLDoff:BY {TIMe|DEFAult| RANDom|AUTO}
TRIGger:A:HOLDoff:BY?
2-566 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual