Command Syntax
Table 2-12: Block Argument (cont.)
Symbol Meaning
<DChar> A character with the hexadecimal equivalent of 00 through FF (0
through 255 decimal)
<Block>
A block of data bytes dened as: <Block> ::=
{#<NZDig><Dig>[<Dig>...][<DChar>...]
|#0[<DChar>...]<terminator>}
<NZDig> s
pecies the number of <Dig> elements that follow. Taken together,
the <NZDig> and <Dig> elements form a decimal integer that species how
many <DChar> elements follow.
NOTE. The digit <NZDig> is in hexadecimal format. This deviates slightly from
the IEEE 488.2 specication that it be in decimal format, as extra allowances
must be made for data lengths that are greater than 999,999,999 <DChar>
elemen
ts (for example 500M record lengths at 2 bytes per point).
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-9