This screen contains the parameters for setting up the threshold values for
errors and Pass/Fail status.
Start Trigger, Stop Trigger
Use the drop-down menu(s) to select the appropriate trigger. Start and stop
triggers include the following:
LOF, OOF, ODU-AIS etc.
Error Free Period
Alarms or Errors set up in stop trigger are not in a measured signal within the
time of this cycle, end the switching time measurement.
Threshold
Allows you to specify the maximum duration of the selected trigger(s). Valid
values are from 0.000 ms to 10000.000 ms.
7.2.3 Test Results
7.2.3.1 Summary
Touching the Summary button in the navigation area will display the screen
shown below.
The following information is displayed:
Start Trigger
Stop Trigger
Average switching time : Average time of switching
Max. Time : Maximum time of switching
Threshold : The threshold value used in the test
7.2.3.2 Detailed
Touching the Detailed button in the navigation area will display the screen
shown below.