Description of device parameters Proline Promag 300 Modbus RS485
208 Endress+Hauser
Description Use this function to switch simulation of the switch output on and off. The display
alternates between the measured value and a diagnostic message of the "Function check"
category (C) while simulation is in progress.
Selection Off
On
Factory setting Off
Additional information Description
The desired simulation value is defined in the Switch status 1 to n parameter.
Selection
Off
Switch simulation is switched off. The device is in normal measuring mode or another
process variable is being simulated.
On
Switch simulation is active.
Switch status 1 to n
Navigation  Expert → Diagnostics → Simulation → Switch status 1 to n (0463–1 to n)
Description Use this function to select a switch value for the simulation. In this way, users can verify
the correct adjustment of the switch output and the correct function of downstream
switching units.
Selection Open
Closed
Additional information Selection
Open
Switch simulation is switched off. The device is in normal measuring mode or another
process variable is being simulated.
Closed
Switch simulation is active.
Relay out. 1 to n sim
Navigation  Expert → Diagnostics → Simulation → Relay out. 1 to n sim (0802–1 to n)
Description Use this function to switch simulation of the relay output on and off. The display
alternates between the measured value and a diagnostic message of the "Function check"
category (C) while simulation is in progress.
Selection Off
On
Factory setting Off