Proline Promag 300 Modbus RS485 Description of device parameters
Endress+Hauser 209
Additional information Description
The desired simulation value is defined in the Switch status 1 to n parameter.
Selection
Off
Relay simulation is switched off. The device is in normal measuring mode or another
process variable is being simulated.
On
Relay simulation is active.
Switch status 1 to n
Navigation  Expert → Diagnostics → Simulation → Switch status 1 to n (0803–1 to n)
Prerequisite The On option is selected in the Switch sim. 1 to n parameter parameter.
Description Use this function to select a relay value for the simulation. In this way, users can verify the
correct adjustment of the relay output and the correct function of downstream switching
units.
Selection Open
Closed
Additional information Selection
Open
Relay simulation is switched off. The device is in normal measuring mode or another
process variable is being simulated.
Closed
Relay simulation is active.
Puls.outp.sim.
Navigation  Expert → Diagnostics → Simulation → Puls.outp.sim. (0988)
Description Use this function to switch simulation of the double pulse output on and off. The display
alternates between the measured value and a diagnostic message of the "Function check"
category (C) while simulation is in progress.
Selection Off
Fixed value
Down-count. val.
Factory setting Off