Proline Promag 300 Modbus RS485 Description of device parameters
Endress+Hauser 49
Description Displays the current totalizer overflow.
User interface Integer with sign
Additional information Description
If the current totalizer reading exceeds 7 digits, which is the maximum value range that
can be displayed by the operating tool, the value above this range is output as an overflow.
The current totalizer value is therefore the sum of the overflow value and the totalizer
value from the Totalizer val. 1 to n parameter.
User interface
The unit of the selected process variable is specified for the totalizer in the Unit
totalizer parameter (→  168).
Example
Calculation of the current totalizer reading when the value exceeds the 7-digit display
range of the operating tool:
Value in the Totalizer val. 1 parameter: 1 968 457 m
3
Value in the Tot. overflow 1 parameter: 2 ⋅ 10
7
(2 overflows) = 20 000 000 [m
3
]
Current totalizer reading: 21 968 457 m
3
"Input values" submenu
Navigation  Expert → Sensor → Measured val. → Input values
Input values
Current input 1 to n
→  49
Val.stat.inp. 1 to n
→  50
"Current input 1 to n" submenu
Navigation  Expert → Sensor → Measured val. → Input values → Current
input 1 to n
Current input 1 to n
Measured val. 1 to n (1603–1 to n)
→  50
Measur. curr. 1 to n (1604–1 to n)
→  50