Chapter 7 - Gantry Page 237
5.) Refer to Figure 7-13. Replace and lay the cables inside the retractor. The length of each cable
that protrudes from the side to which the bracket is assembled is given in Table 7-2.
Table 7-2: Cable Connections to Retractor Bracket
Note: Connect the cables so that their minimum bending radius is 12.5 mm.
6.) Tie-wrap the bundle of cables on the bracket side to prevent the cables from moving.
7.) Close the retractor by fastening the clips.
Cable Part Number Connection
Length of Cable
Protruding from
Retractor
Power 46-311772P1 TS1 170 cm
Power 46-311772P1 TS2 170 cm
Power 46-311772P1 TS3 170 cm
RS232 46-311828G1 + G2 J13 85 cm
Ethernet 46-311844G2 J11 85 cm
Fiber-Optic
(Routed inside conduit.)
46-311920P3 J15 70 cm
RS485 46-311629P1 J8 207 cm
RS485 46-311629P1 J16 73 cm
New Ground 2295967 PET Left Leg 92 cm
E-Stop/Sources CT-PET 2295382 J101 85 cm
New Ethernet 295968 J102 85 cm