Page 70 Chapter 2 - Discovery LS System Safety Information
CAUTION
HIGH CURRENT
LOCK-OUT AND TAG-OUT
ELECTRONICS CABINET AT
THE PDU BEFORE ANY
SERVICE IN THIS AREA
Yellow
Location:
Electronics Cabinet: At the 5 volt high current
circuit, from the power supply to the VME
backplane.
Instructions:
A high current hazard is present at the cabinet
power supply and the VME backplane. Follow the
label instructions, or severe burns may be caused
by accidental short circuits.
USE SAME TYPE AND SAME
RATING REPLACEMENT FUSES
White
Location:
Fused areas.
Instructions:
The fuses in the PCU must be replaced with the
same type and rating fuses in order to provide the
same overcurrent protection.