List and sweep mode
R&S
®
SMA100B
200User Manual 1178.3834.02 ─ 09
You can define up to 10 markers, and assign one of them to the output at a time.
When the sweep run has reached the marker frequency, the R&S SMA100B acti-
vates the marker signal and supplis it at the Marker User1 connector. The duration
of the active signal corresponds to the dwell time of a step.
Settings
Frequency x................................................................................................................ 200
State............................................................................................................................200
Marker Polarity............................................................................................................201
Active Marker.............................................................................................................. 201
Frequency x
Sets the frequency that the selected marker indicates.
When the specified value is used, the polarity of the marker signal changes, as
selected with the parameter .
Remote command:
[:SOURce<hw>]:SWEep[:FREQuency]:MARKer<ch>:FREQuency on page 711
State
Enables the selected marker.
A marker signal at the output connector marks the moment the specified frequency val-
ues is used.
Remote command:
[:SOURce<hw>]:SWEep[:FREQuency]:MARKer<ch>:FSTate on page 711
Sweep mode settings