Remote control commands
R&S
®
SMA100B
517User Manual 1178.3834.02 ─ 09
Example:
SENS:SWE:HCOP:DEV:LANG CSV
selects output format *.csv.
SENS:SWE:HCOP:DEV:LANG:CSV:ORI HOR
selects horizontal orientation
SENS:SWE:HCOP:DEV:LANG:CSV:SEP SEM
selects ";" as the separator between the values
SENS:SWE:HCOP:DEV:LANG:CSV:DPO DOT
selects "." as decimal point
SENS:SWE:HCOP:DATA?
queries the measurement data of the current traces
Response:
#2651009500000;1019000000;1028500000;1038000000
-9.5;-9.7;-6.3;-2.5
The hash symbol # introduces the data block. The next number
indicates how many of the following digits describe the length of
the data block. In the example, the 2 following digit indicates the
length to be 65 characters.
Because horizontal representation is selected, a row with all the
x-values of the active trace (frequency) follows. The second row
contains all the y-values of the active trace (power). The rows
end with a new line (each counts as one character).
Note: if more than one trace is active, the third row contains the
x values of the second active trace, and so on.
Example:
SENS:SWE:HCOP:DEV:LANG:CSV:ORI VERT
selects horizontal orientation
SENS:SWE:HCOP:DATA?
queries the measurement data of the current traces
Response:
#2681009500000;-9.5; 1019000000;-9.7; 1028500000;-6.3;
1038000000;-2.5;
for vertical representation the length of the data block is 68
the first power value, the second row contains the second fre-
quency value of the active trace followed by the second power
value, and so on. The rows end with a new line (each counts as
one character).
Note: if more than one trace is active, the first row also contains
the value pairs of the second active trace, and so on.
Usage:
Query only
:SENSe[:POWer]:SWEep:HCOPy:DEVice <Device>
Defines the output device. The setting is fixed to FILE, i.e. the hardcopy is stored in a
file.
Parameters:
<Device> FILE | PRINter
*RST: FILE
Power sensor measurement subsystems