Remote control commands
R&S
®
SMA100B
526User Manual 1178.3834.02 ─ 09
Parameters:
<YValues> string
Example:
SENSe:POW:SWE:TIME:REF:DATA:YVAL -30,10
Sets the y value of reference "Point A" to -30 dBm, and the
value of "Point B" to 10 dBm.
Manual operation: See "Pow (Y)" on page 254
:SENSe[:POWer]:SWEep:POWer:RMODe <RMode>
Selects single or continuous mode for measurement mode power in power analysis.
Parameters:
<RMode> SINGle | CONTinuous
*RST: CONTinuous
Example:
SENS:SWE:POW:RMOD SING
Selects single measurement.
Manual operation: See "Execution - Power Analysis" on page 258
:SENSe[:POWer]:SWEep:POWer:SPACing[:MODE] <Mode>
Selects the spacing for the frequency power analysis.
Parameters:
<Mode> LINear
*RST: LINear
Example:
SENS:SWE:FREQ:SPAC:MODE LIN
Sets linear spacing of the sweep.
Manual operation: See "Spacing - Power Analysis" on page 258
:SENSe[:POWer]:SWEep:POWer:STARt <Start>
Sets the start level for the power versus power measurement.
Parameters:
<Start> float
Range: -145 to 20
Increment: 0.01
*RST: 1MHZ
Example:
SENS:SWE:POW:STAR -20DBM
Sets the start level to -20 dBm
Manual operation: See "Min - Power Analysis" on page 256
:SENSe[:POWer]:SWEep:POWer:STEPs <Steps>
Sets the number of measurement steps for the power versus power measurement.
Power sensor measurement subsystems