Remote control commands
R&S
®
SMA100B
553User Manual 1178.3834.02 ─ 09
Parameters:
<Power> DBM | DBUV | WATT
*RST: DBM
Example:
:SENS2:UNIT DBM
Selects dBm as unit for the measured value returned by com-
mand READ.
:READ2?
Response: 7.34
7.34 dBm are measured by sensor 2.
Manual operation: See "Level (Peak) / Level (Average)" on page 241
:SENSe<ch>[:POWer]:APERture:DEFault:STATe
<UseDefAp>
Deactivates the default aperture time of the respective sensor.
To specify a user-defined value, use the command :SENSe<ch>[:POWer]:
APERture:TIMe on page 553.
Parameters:
<UseDefAp> 0 | 1 | OFF | ON
*RST: 1
Example: See Example "Performing a power measurement with a fixed fil-
ter" on page 548.
Manual operation: See " Default Aperture Time" on page 244
:SENSe<ch>[:POWer]:APERture:TIMe <ApTime>
Defines the aperture time (size of the acquisition interval) for the corresponding sensor.
Parameters:
<ApTime> float
Range: depends on connected power sensor
Increment: 1E-9
*RST: depends on connected power sensor
Example: See Example "Performing a power measurement with a fixed fil-
ter" on page 548.
Manual operation: See "Aperture Time" on page 244
:SENSe<ch>[:POWer]:CORRection:SPDevice:SELect <Select>
Several S-parameter tables can be stored in a sensor. The command selects a loaded
data set for S-parameter correction for the corresponding sensor.
Parameters:
<Select> float
*RST: 0
SENSe, READ, INITiate and SLISt subsystems