548User Manual 1178.3834.02 ─ 09
Example: Performing a power measurement with a fixed filter
Prerequisite: The sensor is connected to the instrument and mapped to the first sensor
channel.
SENSe1:SOURce RF
// Sensor measures the power of the RF signal.
SENSe1:FILTer:TYPE NSRatio
// Selects fixed noise filter mode.
SENSe1:FILTer:NSRatio 0.02 DB
// Sets the maximum noise component in the result to 0.02 DB.
SENSe1:FILTer:NSRatio:MTIMe 10
//Limits the settling time to 10 seconds.
SENSe1:APERture:DEFault:STATe 0
// Deactivates the default aperture time of the sensor.
SENSe1:APERture:TIMe 10e-6
// Sets the aperture time to 10 us.
SENSe1:UNIT DBM
// Selects unit dBm for the measured value.
INITiate:CONTinuous ON
// Switches the continous power measurement on.
READ?
// Triggers the measurement and displays the results.
:SLISt[:LIST]?................................................................................................................ 549
:SLISt:SCAN[:STATe]......................................................................................................549
:SLISt:SCAN:LSENsor....................................................................................................549
:SLISt:SCAN:USENsor................................................................................................... 550
:SLISt:CLEar:LAN.......................................................................................................... 550
:SLISt:CLEar:USB..........................................................................................................550
:SLISt:CLEar[:ALL].........................................................................................................551
:SLISt:ELEMent<ch>:MAPPing........................................................................................551
:SLISt:SENSor:MAP.......................................................................................................551
:INITiate<hw>[:POWer]:CONTinuous............................................................................... 551
:READ<ch>[:POWer]?....................................................................................................552
:SENSe<ch>:UNIT[:POWer]............................................................................................552
:SENSe<ch>[:POWer]:APERture:DEFault:STATe.............................................................. 553
:SENSe<ch>[:POWer]:APERture:TIMe.............................................................................553
:SENSe<ch>[:POWer]:CORRection:SPDevice:SELect.......................................................553
:SENSe<ch>[:POWer]:CORRection:SPDevice:STATe........................................................554
:SENSe<ch>[:POWer]:CORRection:SPDevice:LIST?.........................................................554
:SENSe<ch>[:POWer]:DISPlay:PERManent:PRIority.........................................................554
:SENSe<ch>[:POWer]:DISPlay:PERManent:STATe........................................................... 554
:SENSe<ch>[:POWer]:FILTer:LENGth:AUTO?.................................................................. 555
SENSe, READ, INITiate and SLISt subsystems

Table of Contents

Other manuals for R&S SMA100B

Related product manuals